(7) Press the NORMAL SCREEN Control Key. 

s. Perform Low Flow adjustment as follows: 

(1) Connect Flow Meter to ETC02 Patient Port on the monitor front panel. 

(2) Adjust AS191 R7 for areading of 150 ml ±5 ml. 

t. Set the ETC02 to High Flow as follows: 

(1) Turn the Rotary Knob to highlight the ETCG2 icon. 

(2) Press the Rotary Knob to select the ETG02 menu. 

(3) Turn the Rotary Knob to highlight the FLOW menu selection. 

(4) Press the Rotary Knob to select FLOW. 

(5) Turn the Rotary Knob to highlight the HIGH menu selection. 

(6) Press the Rotary Knob to select HIGH. 

(7) Press the NORMAL SCREEN Control Key. 

u. Perform High Flow adjustment by adjusting AS191 R4 for a reading of230 ml ±5 
ml. 

v. Repeat substeps r, s, t and u as necessary. 

w. Disconnect the Flow Meter from the ETC02 port on the monitor front panel. 

x. Perform ETC02 Verification (Section II). 

4.5 Respiration Adjustment Procedure. Perform the following procedure to adjust the 
Respiration System. 

a. Remove top cover and front panel assembly (Reference Section IE: Disassembly 
Procedures). 

b. Turn on the monitor. 

c. Press the SETUP Control Key. 

d. Turn the Rotary Knob to highlight the DISPLAY SETUP menu selection. 

e. Press the Rotary Knob to select DISHAY SETUP. 

f. Turn the Rotary Knob to highlight the TRACE C menu selection. 

g. Press the Rotary Knob to select TRACE C. 

h. Turn the Rotary Knob to highlight the RESP menu selection. 

i. Press the Rotary Knob to select RESP. 

j. Press the NORMAL SCREEN Control Key. 

k. Turn the Rotary Knob to highlight the RESP window. 

l. Press the Rotary Knob to select RESP. 

m. Turn the Rotary Knob to highlight the SIZE menu selection. 

n. Press the Rotary Knob to select SIZE. 


4-9 



o. Turn the Rotary Knob to highlight the 10 Ohm menu selection. 

p. Press the Rotary Knob to select 10 Ohm 

q. Press the NORMAL SCREEN Control Key. 

r. . v Set the Patient Simulator Respiration output to 3 Ohms at a rate of 30BPM with the 

7 Baseline Impedance set to 500 Ohms. 


s. Disconnect the RA (white) and LA (black) leads from the Patient Simulator. 


t. 

Cu) 

u. 


Using a DMM, set to measure "TRUE RMS AC CURRENT' with the range set to 
measure 90 uA AC (100 Ohm input impedance), connect the meter to the RA and 
LA leads and adjust AB99 R25 for 90 pA AC (±5pA). gima SoA. em 

Reconnect leads to the patient simulator. ' - 


v. 


w. 


x. 


y- 


Connect a X10 scope probe to AB99 TP1 using AB99A GND as reference ground. 
Set the scope controls to 50mv DC input coupling, 5ps/cm 


Adjust AB99 Cl 5 and C22 for a 
minimum waveform (approximately) 
lVp-p (2 divisions) (See Figure 4-2) 
with equal negative peaks and correct 
phasing representing the positive 
portion of a sinusoidal waveform. 
Recheck steps s through w. 



Figure 4*2. Respiration Waveform 


Disconnect scope probe from TP 1 and connect it to TP2. 


z. Set the scope controls to 0.2V and input switch to ground. Adjust the scope baseline 
for 0VDC then change the switch back to DC input coupling. 

aa. Adjust AB99 R23 for 6Vdc (3 divisions). 

ab. Disconnect scope probe from TP2 and hold it on to AB99 P4 pinl. 

ac. Set the scope controls to 50mv DC input coupling. 

ad. Adjust AB99 R18 for 2.14 Vp-p (approximately 4.3 divisions). 

ae. Remove scope probe. 

af. Verify waveform on the display. 

ag. Perform RESP Calibration Verification (Section II). 


4-10 












SECTION V 

TROUBLESHOOTING FLOW CHARTS 

Table of Contents 


Failure Indication Page Number 

Monitor Will Not Power Up on Facility Line Power. 5-2 

Power Supply Test. 5-3 

Won't Power Up on Battery Power . 5-4 

No Display Backlight.. 5-5 


List of Symbols 


See 

I Power Supply 1 
VLoading Chart/ 


Round or Oval: These symbols are 
used to provide information. 


[ Check AC Power 
Connections. 


Square or Rectangle: These symbols are used 
when technician actions are required. 



Diamond: This symbol is used when there 
is a decision for the technician to make. 



Circled Letter: This symbol is used to move 
the technician from place to place on the 
Troubleshooting Charts. 


Millennia™ Troubleshooting Flow Charts - Page 5-1: Table of Contents and List of Symbols 







Monitor will not 
s_Power Up.^, 


© 


Check AC Power 
Connections. 

Line In 
Good 


Line In 
Bid 


Correct 

Problem. 


/ |s \ 

/ the Line \ 

Voltage Selection 

v(on back panel)/ 

X correct? / 


Setting 
, r Correct 

/l\ 
A/C ChgX 
Indicator/ 

X0n?X 


Setting Reset to correct 
incorrect Line Voltage Setting. 


Check Main Fuse 
AC Fuses, fod 


Replace 

Fuses. 


Perform System Test and Verification, before 
returning to operation. For Troubleshooting 
Assistance, contact Invivo Research Technical 
Service at (407) 275-3220 or (800) 331-3220; 
FAX at (407) 249-2022. 

NOTE 

Invivo Research, Incorporated 
assumes no responsibility for 
damages incurred during, or as a 
result of, maintenance by Technicians 
who are not certified by Invivo 
Research, Incorporated. 


r WARNING 

As with all electronic equipment, 
potentially harmful voltages exist in live 
circuitry. Due care and caution, with 
strict adherence to electrical safety 
practices, must be exercised every time 
the monitor is connected to line power 
.with any part of its casing removed. A 


Using a DMM, measure 
for 24VDC +/-3VDC across Voltage Out 
the 10,000uf capacitor of Tolerance 
on the AC264. 

Voltage within 
Tolerance 


Check AC Power Switch, Rectifier and 
Transformer for proper connections and 
operation. Correct or Replace as necessary. 


Check Front Panel 
Power Switch. 



Replace Front Panel 
Power Switch. 


Correct 

Discrepancies. 



Check 

Good 

V 

Perform ^ 
Power Supply 
^ Test ^ 


Check all internal cables or 
cables and connections.! connections bad 

Cables and 
Connections Good 


Correct 

Discrepancies. 


^ Contact 
Invivo Research, Inc. 
^technical Support^ 


Millennia™ Troubleshooting Flow Charts - Page 5-2: Monitor Will Not Power Up on Facility Line Power 



















Power Up. 



NOTE 

Invivo Research, Incorporated 
assumes no responsibility for 
damages incurred during, or as a 
result of, maintenance by Technicians 
who are not certified by Invivo 
Research, Incorporated. 


r 


WARNING 




_z_ 

Power on then check the AB69A 
for the following voltages using 
J3 P7 as reference ground: 

PI = -5 Vdc (+/-50mv) 

P5 = +12.1 Vdc (+/-0.2 Vdc) 
P6 = 12.1 Vdc (+/- 0.2 Vdc) 

P8 = 24 Vdc (+/- 3 Vdc) 



Yes 


_i_ 

Install the boards one at a 
time to check for loading. 



As with all electronic equipment, 
potentially harmful voltages exist in live 
circuitry. Due care and caution, with 
strict adherence to electrical safety 
practices, must be exercised every time 
the monitor is connected to line power 
L with any part of its casing removed. 



1 

r 

Yes 

r 

Replace 

AB69A. 

* 

¥ 


No 

S_w 

Check Cables, Connections 
and Line Power. 
Correct any discrepancy. 

f -;- 






6 


No 

Replace 


f W 

AB69. 



O 


Perform System Test and Verification, before 
returning to operation. For Troubleshooting 
Assistance, contact Invivo Research Technical 
Service at (407) 275-3220 or (800) 331-3220; 
FAX at (407) 249-2022. 


Millennia™ Troubleshooting Flow Charts - Page 5-3: Power Supply Test 







Service at (407) 275-3220 or (800) 331-3220; 
FAX at (407) 249-2022. 


Millennia™ Troubleshooting Flow Charts - Page 5-4: Won’t Power Up on Battery Power 










Remove AB77 and check 


connectors at the AB09 and 
P7 on the AB77. 

Good 


_i_ 

Install AB77, leaving the 
display disconnected. 


* 


Turn On 
Power. 



Yes 


t 


Replace 

Backlight. 



Bad 

Replace 

W 

Batteries. 


5 




WARNING 




As with all electronic equipment, 

n/vtarvHqHw harmful wrJ+Qrtoo avic* iri liwn 
pui.v/iluuiij iitiimiui vuuciyOo cajol in live 

circuitry. Due care and caution, with 
strict adherence to electrical safety 
practices, must be exercised every time 
the monitor is connected to line power 
L with any part of its casing removed. 


o 

Perform System Test and Verification, before 
returning to operation. For Troubleshooting 
Assistance, contact Invivo Research Technical 
Service at (407) 275-3220 or (800) 331-3220; 
FAX at (407) 249-2022. 



Millennia™ Troubleshooting Flow Charts - Page 5-5: No Display Backlight 














Appendix A 

Assembly and Schematic Drawings 

Part Number Dwg Nomenclature Page Number 

3500 Millennia Interconnect Diagram.A-2 

3500 Mllermia 3500 Vital Signs Monitor Assembly Diagram.A-3 

AB09A LCD Inverter (with Heat Shrink) Assembly Diagram. A-13 

AB65 A/D Assembly Diagram. A-14 

AB65 A/D Schematic Diagram. A-15 

AB66 Patient Front End Assembly Diagram . A-20 

AB66 Patient Front End Schmatic Diagram. A-23 

AB68 Keyboard Interface Assembly Diagram. A-36 

AB68 Keyboard Interface Schematic Diagram. A-37 

AB69 Power Supply Assembly Diagram. A-38 

AB69 Power Supply Schematic Diagram. A-39 

AB69A Power Switcher/Battery Charger Assembly Diagram . A-42 

AB69A Power Switcher/Battery Charger Schematic Diagram. A-43 

AB70 Multi-Function FO Assembly Diagram. A-44 

AB70 Multi-Function I/O Schematic Diagram. A-45 

AB72 Power Interconnect Assembly Diagram. A-49 

AB72 Power Interconnect Schematic Diagram. A-50 

AB74 Invasive Pressures Assembly Diagram. A-51 

AB74 Invasive Pressures Schematic Diagram . A-52 

AB77 PCMCIA/VGA Display Assembly Diagram. A-53 

AB77 PCMCIA/VGA Display Schematic Diagram. A-54 

AB91 ETC02 Pneumatics Controller Assembly Diagram. A-56 

AB91 ETC02 Pneumatics Controller Schematic Diagram. A-57 

AB92 Assy: PCA, Connector Sensor, 3500 . A-59 

AB92 Sch: Patient Connector Sensor, 3500 . A-60 

AB93 Assy: PCA, Rear Panel Connectors, 3500 . A-61 

AB93 Sch: PCA Rear Panel Connectors, 3500 . A-62 

AB97 Assy: PCA, Sharp 321 LCD Display, Cable Adapter, 3500. A-63 

AB97 Sch: PCA Sharp 321 LCD Active Display, Cable Adapter, 3500 . A-64 

AB105 Assy: PCA, Sharp 35PX Passive Cable Adptr. A-65 

AB105 Sch: PCA Sharp 35PX Passive Cable Adptr.. A-66 

AS 137 Assy: Speaker, 3500 . A-67 

AS191 Assy: Bid-Tidal C02 Module . A-68 